Home /
Products
/ Test &
Measuring Instruments
/ Automatic C & Tand Bridge ( Model : MLS-5A & HLS-5A) |
|
Automatic C & Tand Bridge |
|
Model : MLS-5A & HLS-5A ( Front Panel Controlled) |
| |
|

Automatic Capacitance & Tan d Test
Set
This set includes Bridge Model:MLS5A & HV
Supply model :HLS5A , Touch-screen
controlled & necessary cables & very flexible,
user friendly Visual C++ windows based
Software. The use of Touch-screen & Windows
based software makes the system more
flexible easy to handle & accurate Data can be
stored, Retrieved and analyzed at site.
Temp correction for the tan delta of object is
automatically done as per IEC. |
|
FEATURES |
 |
- Automatic Measurement of Capacitance & Tand as per the test plan
by
auto-balancing .No need of manual balancing. Automatic voltage
setting
through software.
- Indication of Leakage current.
- Data logging with graph of voltage vs. Tand to know the tand gradient & data
analysis for comparison with old results / manufacturer's data.
- Temperature correction as per IEC for
Tand.
- Controls on front panel , totally controlled by Touch-screen.
Executes all the test
modes/voltage automatically once fed
- On site printout.
- Suitable for extraneous conditions in 400KV switchyard under heavy induction.
- Unlimited Data-storage in Laptop
|
|
MEASUREMENTS |
 |
The system will provide following Information.
|
PARAMETER |
RANGE |
RESOLUTION |
ACCURACY |
| CAPACITANCE |
1pF – 1.1 uF |
4 Digits |
± 0.2% of the reading
± 1pF ( UST )
± 10pF ( GST ) |
| TAN DELTA |
0 % – 100%
(± infinity) |
0.0001 |
± 1% of the reading ± 0.02% |
VOLTAGE
Measurement |
0 – 5 KV |
1.0 V |
± 1%, ± 1digit |
CURRENT
Measurement |
0 – 2000 mA |
1mA |
± 2 % ± 1digit |
| POWER FACTOR |
0.01% – 100% |
0.00001 |
± 1.0% of the reading ± 0.05% |
| POWER |
0 – 2 KW |
0.1 mW |
± 1 % of reading |
INTERFERANCE
CURRENT |
0 – 100 mA |
1mA |
± 1% of reading |
| LINE FREQUENCY |
45Hz – 65 Hz |
0.01Hz |
± 0.02Hz |
Analysis Software :
Windows based analysis software for Tand Vs Time, Tan
d Vs Voltage.
Temperature Range : 100C to 500C.
( Conforms to Dry Heat Test 500C, Damp Heat
Test 400C)
Humidity Range : Ambient to 95 %
With built in 100 pF / 200pF capacitor.
Standard Capacitor : Standard Capacitor is in built into the HV power
supply.
Capacitance : 100 pf ± 0.5%, Actual value is indicated on capacitor.
Tand : Less than 0.0002
Test Voltage : Up to 5 KV RMS
Max. Voltage (1 Minute) : 5 KV RMS
|
|
TECHNICAL SPECIFICATIONS FOR HIGH VOLTAGE POWER SUPPLY |
 |
Test Voltage 0 to 5 KV
Rated Current 120mA Continuous & 240mA intermittent.
Rated Power 600VA & 1200VA intermittent .
Output Frequency 50 Hz ( Lines Frequency )
Automatic balancing through windows based software & touch-screen.
No need of manual voltage control.
Automatic selection of seven different modes of operation for measurement of 3
terminal object from windows based software .
1) UST - Ungrounded Specimen test - 3 combinations
This mode is used when the object under test is not grounded. This test
provides most accurate results.
2) GST - Grounded Specimen test
This mode is used when the object under test is permanently grounded. This
test is used more often in outdoor installations, power systems etc.
3) GSTg - Grounded specimen test with guard. 3 combinations.
This mode is used for measuring them stray capacitances and separating
them from basic measurement in GST mode.
Indication : 3 ½ Digital Panel Meter.
Standard capacitor : 100pF/200pF 5KV built-in.
Input Frequency : From 50 Hz ( lines frequency) , ± 5%.
Power Required : 230 VAC ± 10%, 1 phase AC , 50Hz.
Safety Features :
· Zero start control
· Open ground indicating lamp with double grounded connection
· External interlock.
· HT cut off on overload
· HT ON & Supply ON indication.
· All operating controls at earth potential
· Necessary terminals & Sockets for connection to Bridge |
|
|
| To keep pace with the latest technology
Sivananda Electronics reserves the right to update it's system.
Hence, information in the brochure is likely to change without
notice. |
|
|
|