These instruments have been specially designed for routine
testing and sorting of great
number of Paper and Polypropylene Capacitors used for correcting Loss Factor of
industrial loads. Tan d is read directly in terms of
percentage. The instrument has an essential advantage over the customary bridges
as measurements are carried out under actual working conditions (50 Hz, 150
Volts or high voltage upto 11KV) and that no tuning is necessary prior to every
single reading. This results in higher sorting speed, hence in more economical
productive tests. No exacting demands are made on the alertness of the checker.
The measured value can be read directly on a meter . The
Loss Factor Meter makes use of the fact that the current through the capacitor
will not lead exactly 90 from the voltage across it. The extent of deviation in
the phase is measured with an ultra sensitive phase detector comprising of
solid-state components only. The capacitor is coupled to the circuit by a low
loss network. A separate meter for calibration is provided to compensate
been made to use the best possible components to ensure electrical stability.
The measurement is done at AC voltage between 150 V and 11 KV, 50 Hz. The
instrument measures the Loss Factor as low as 0.0002 and the indication is
directly given on the Digital Panel Meter. The capacitance range is 0.1 mfd to
1000 mfd. The capacitors can be checked at the rate of 200 per hour.
Isolating transformer of adequate rating is required depending on the size of
the biggest capacitor to be tested. Speed of testing can be increased
substantially if the equipment is used in conjunction with "Control Unit" Model,
LPO - 5D/LPO - 5MD can be used in conjunction with Capacitance Meter LSO - 9D
for simultaneous indication of Value and Loss Angle.The equipment is available
in two models as per the specifications given overleaf.
|
TECHNICAL SPECIFICATIONS |
 |
| MODEL |
LPO - 5D |
LPO - 5MD |
| Capacitance Range |
1 to 1000 mfd |
1 to 30 mfd |
| Loss Factor Range |
0.0002 to 0.1999 |
0.0002 to 0.1999 |
| Resolution |
0.0001 tan d |
0.0001 tan d |
A. C. Voltage applicable
by external power source |
150 to 600 V |
150 to 600 V |
| Indication |
3½ digit DPM |
3½ digit DPM |
| Speed of testing 200 capacitors per hour |
with a suitable jig 200 capacitors per hour |
with a suitable jig |
| Warm up time |
5 minutes |
5 minutes |
| Weight |
11 Kg. |
11 Kg. |
| Construction for bench use |
Sturdy metal cabinet
for bench use |
Sturdy metal cabinet |
| Dimensions |
435 x 345 x 192 mm |
420 x 270 x 225 mm |
|
|