Test Voltage :
0 to 12 KV / 50 Hz
Rated Power :
1 KVA Maximum , 100mA
Long Term , Current Capacity
This power capacity is adequate for capacitance upto
0.0mF at 1 KV
(The H.V. power supply with higher power capacity upto 15 KVA or with a tap
at lower voltage for higher current is also available).
Switch for three different types
measurement
1) UST - Ungrounded Specimen test - 3 combinations
This mode is used when the object under test is not grounded. This test
provides most accurate results.
2) GST - Grounded Specimen test
This mode is used when the object under test is permanently grounded. This
test is used more often in outdoor installations, power systems etc.
3) GSTg - Grounded specimen test with guard. 3 combinations.
This mode is used for measuring the stray capacitances and separating
it from basic measurement in GST mode.
Indication : Two 3½ digit digital panel meters one for voltage indication
with 19.99 KV full scale and other for current indication.
Safety Features :
- Zero start control
- Open ground indication lamp with double grounded connection External
interlock
- External Interlock
- HT Cut off on overload
- HT ON & Supply ON indication
- All operating controls at earth potential.
- Necessary terminals & sockets. For connection to MLS-11D M
Power Required : 230 VAC ± 10%, 50 Hz
Size : 510 x 550 x 590 mm
Standard Capacitor : Standard Capacitor is compressed SF6 gas filled
type and is built into HV power supply.
Capacitance : 100 pF ± 5%, actual value is
indicated on capacitor.
Tan : Less than 0.00001
Dielectric : SF6
Test Voltage : upto 12 KV RMS.
Max. Voltage (1 minute) : 12 KV RMS. |